A new pyruvate oxidase biosensor based on 3-mercaptopropionic acid/6-aminocaproic acid modified gold electrode

Makaleler > Tam Makale · 2013

Yazarlar: Ezgi BAYRAM, Erol AKYILMAZ

Artificial Cells, Nanomedicine, and Biotechnology, SCI Expanded İndekslerine Giren Dergi, Uluslararası, Scopus, 2013, cilt 42, ss 418-422

DOI: 10.3109/21691401.2013.815626

Özet

In the biosensor construction, 3-mercaptopropionic acid (3-MPA) and 6-aminocaproic acid (6-ACA) were used for forming self-assembled monolayer (SAM) on a gold disc electrode and pyruvate oxidase was immobilized on the modified electrode surface by using glutaraldehyde. Biosensor response is linearly related to pyruvate concentration at 2.5-50 μM, detection limit is 1.87 μM and response time of the biosensor is 6 s for differential pulse voltammograms. From the repeatability studies (n = 6) for 30.0 μM pyruvate revealed that the average value ([Formula: see text]), standard deviation (S.D) and coefficient of variation (CV %) were calculated to be 31.02 μM, ± 0.1914 μM and 0.62%, respectively.

Detay görüntüle