Investigation of dose distribution of TheraSeed 200 103Pd brachytherapy seed on coaxial cylindrical phantom shells using Monte Carlo simulation
Articles > Article · 2011
Authors: Berkay CAMGÖZ, Mehmet Nurullah KUMRU
Annals of Nuclear Medicine, Journal Indexed in SCI Expanded, International, WoS, 2011, vol 25, pp 371-379
DOI: 10.1007/s12149-011-0472-4
Summary
In this study, radial dosimetric characterization for specific internal geometry of seed source was determined. In this study, the separate contribution of radioactive components of the seed's unlikely single radioactive rod on dose distribution was investigated using model seed TheraSeed 200 source at parallel shells in a water phantom.