Is an automatic check of the anesthesia machine sufficient before anesthesia induction? A structural obstruction in the breathing circuit

Makaleler > Tam Makale · 2026

Yazarlar: Cengiz ŞAHUTOĞLU, Gökçe Su SAYGILI, Asuman SARGIN, Canan BOR

Perioperative Care and Operating Room Management, ESCI İndekslerine Giren Dergi, Uluslararası, Scopus, 2026, cilt 43, ss 100653-100653

DOI: 10.1016/j.pcorm.2026.100653

Özet

<p>Breathing&nbsp;circuits&nbsp;most&nbsp;often&nbsp;fail&nbsp;as&nbsp;a&nbsp;result&nbsp;of&nbsp;leaks,&nbsp;but&nbsp;obstructions&nbsp;in&nbsp;the&nbsp;expiratory&nbsp;limb&nbsp;are&nbsp;rare&nbsp;and&nbsp;poorly&nbsp;understood.&nbsp;In&nbsp;this&nbsp;report,&nbsp;we&nbsp;aimed&nbsp;to&nbsp;present&nbsp;unexpected&nbsp;mechanical&nbsp;ventilation&nbsp;difficulties&nbsp;that&nbsp;occurred&nbsp;in&nbsp;pediatric&nbsp;patients.&nbsp;In&nbsp;both&nbsp;cases,&nbsp;airway&nbsp;pressures&nbsp;increased&nbsp;to&nbsp;peak&nbsp;inspiratory&nbsp;pressure&nbsp;(PIP)&nbsp;of&nbsp;40&nbsp;cmH₂O&nbsp;and&nbsp;positive&nbsp;end&nbsp;expiratory&nbsp;pressure&nbsp;(PEEP)&nbsp;of&nbsp;34&nbsp;cmH₂O&nbsp;during&nbsp;both&nbsp;manual&nbsp;and&nbsp;automatic&nbsp;ventilation.&nbsp;Exami-nations&nbsp;of&nbsp;the&nbsp;breathing&nbsp;circuits&nbsp;revealed&nbsp;the&nbsp;presence&nbsp;of&nbsp;partial&nbsp;or&nbsp;total&nbsp;obstructions&nbsp;in&nbsp;the&nbsp;expiratory&nbsp;limb.&nbsp;These&nbsp;obstructions&nbsp;could&nbsp;not&nbsp;be&nbsp;detected&nbsp;by&nbsp;the&nbsp;anesthesia&nbsp;machine,&nbsp;either&nbsp;automatic&nbsp;or&nbsp;circuit&nbsp;check&nbsp;test.</p>