Multiwalled Carbon Nanotubes‐Chitosan Modified Single‐Use Biosensors for Electrochemical Monitoring of Drug‐DNA Interactions

Makaleler > Tam Makale · 2015

Yazarlar: Ceren SENGİZ, Gulsah CONGUR, Ece EKŞİN, Kadriye Arzum ERDEM GÜRSAN

Electroanalysis, SCI Expanded İndekslerine Giren Dergi, Uluslararası, Wos, 2015, cilt 27, ss 1855-1863

DOI: 10.1002/elan.201500107

Özet

Abstract A multiwalled carbon nanotubes (CNT)‐chitosan (CHIT) modified pencil graphite electrode (CNT‐CHIT/PGE) was developed for the first time herein for electrochemical monitoring of the interaction of an anticancer drug, mitomycin C (MC) and DNA. The characterization of unmodified PGE, CHIT/PGE, CNT/PGE and CHIT‐CNT/PGE were performed by scanning electron microscopy and cyclic voltammetry techniques. The oxidation signals of MC and guanine were measured before and after interaction at the surface of CNT‐CHIT/PGEs using differential pulse voltammetry. Electrochemical impedance spectroscopy technique was also successfully utilized for monitoring of the interaction process at the surface of CNT‐CHIT/PGEs in different interaction times.