Explainable AI molecularly imprinted polymer electrochemiluminescence biosensor for discriminative methamphetamine and amphetamine detection
Makaleler > Tam Makale · 2026
Yazarlar: Emre DOKUZPARMAK, Emine SEZER, İrem Nur Ceylan, Nur Ceylin Çetin, Hilal Özçelik, Sinan Akgöl
iScience, Uluslararası, Wos, 2026, cilt 29, ss 117483-117483
DOI: 10.1016/j.isci.2026.117483
Özet
Independent forensic quantification of methamphetamine (MA) and amphetamine (AMP) remains challenging due to their near-identical structures and overlapping electrochemical signatures. We report an end-to-end AI-integrated MIP-ECL biosensing framework employing target-specific MIP-NP/MWCNT/Nafion/Ir(pq) 2 (acac)/SPCE platforms. The Ir(III) luminophore within an MWCNT/Nafion matrix produced markedly superior ECL emission over the [Ru(bpy) 3 ] 2+ benchmark (40 mM DBAE, pH 8.0). MIP nanoparticles conferred high selectivity against structurally related interferents, validated in saliva, urine, and blood plasma, achieving detection limits of 0.11 ng mL −1 for MA and 0.28 ng mL −1 for AMP. A structured explainable machine learning pipeline employing CatBoost with Group-K-Fold cross-validation eliminated data leakage and achieved test R 2 values of 0.9864 (MA) and 0.9675 (AMP). SHAP analysis confirmed that model predictions were consistent with known electrochemical kinetics. This modular framework establishes a generalizable paradigm for intelligent MIP-ECL biosensing with direct applicability in forensic toxicology and point-of-care diagnostics.