The usage of a bismuth film electrode as transducer in glucose biosensing

Makaleler > Tam Makale · 2007

Yazarlar: Uelkue Anik, Suna TİMUR, Meliha Cubukcu, Arben MERKOÇİ

MICROCHIMICA ACTA, SCI Expanded İndekslerine Giren Dergi, Uluslararası, 2007, cilt 160, ss 269-273

DOI: 10.1007/s00604-007-0868-y

Özet

A second generation glucose biosensor was developed by using neutral red (NR) as a mediator and a bismuth film electrode (BiFE) as a transducer along with immobilized glucose oxidase. The linear range was between 0.2 and 2.5mM, and a correlation coefficient of 0.999 was obtained with this electrode. The standard deviation (at 1mM glucose for n = 4) and the coefficient of variation were calculated as +/- 8.07 mu M and 3.4%, respectively. The biosensor was used for the determination of glucose in wine samples.