A method for coincidence timing resolution enhancement

Makaleler > Tam Makale · 2016

Yazarlar: E. E. Ermis, Cüneyt ÇELİKTAŞ, E. Pilicer

Review of Scientific Instruments, Uluslararası, Wos, 2016, cilt 87, ss 053504-053504

DOI: 10.1063/1.4948925

Özet

A method including the coincidence time resolution improvement for a TOF/positron emission tomography system was suggested. The spectrometer for this aim was composed of two NaI(Tl) and two plastic scintillation detectors. Experimental results were supported by FLUKA Monte Carlo simulation program by constructing the detector setup in software medium. Present experimental results verified our previous results and conclusions obtained from the suggested method. It was concluded that better resolutions would help the improvement not only on the TOF gain but also on the spatial resolution, leading to better images and helping the Physician in his/her diagnosis and treatment.

Detay görüntüle